Preparing a qNMR sample so the result holds up
Quantitative NMR is unforgiving in a specific way: almost every error is introduced before the spectrometer is involved.
On this page
Weighing is half the measurement
The result is a ratio of moles, so both masses enter the calculation with equal weight. A balance that reads to five decimal places in grams is not enough if the sample is weighed to two. Static is a real problem with dry powders in a dry room, and a hygroscopic solid gains mass while it sits on the pan. Weigh into a tared vessel, close it, and record the mass immediately.
Impurities standards often arrive in small quantities. The minimum weight that can be weighed without losing accuracy is therefore a design constraint on the whole experiment, not an afterthought: choose the sample size, the internal standard amount and the final volume together so that both masses land well inside the balance's working range.
Solvent, internal standard, and one clean signal
The solvent has to dissolve both materials completely and must not have residual proton signals where the analyte or the standard resonates. Deuterated dimethyl sulfoxide, water and chloroform cover most cases; each brings its own water peak and its own shift for exchangeable protons.
For the internal standard, a single sharp resonance is worth more than a familiar name. It should be chemically inert towards the analyte, non-volatile, and not hygroscopic. Maleic acid, dimethylformamide and trimethylsilylpropanoate are common choices for different solvents, and each fails in a different way when handled carelessly.
Parameters that have to be set on purpose
Quantitative conditions differ from structural ones. Use a full pulse angle or a calibrated one, set a relaxation delay long enough for both signals to recover, and acquire enough scans for the signal-to-noise ratio the specification needs. Keep the probe temperature stable, because integrals drift with shimming and with temperature. Spinning is usually switched off: it improves line shape on old probes and degrades it on gradients.
Integrate whole multiplets, define the same integration width for every spectrum in the series, and process all spectra with identical parameters. An integral ratio computed from spectra processed differently is not a ratio of anything.
What we can supply
Items with an assigned content value on the certificate, the method that produced it, and the reference material used for the assignment. Where the assignment is by quantitative NMR the certificate names the internal standard, and the value is stated on a defined basis so you can convert it to the material you weigh. Solutions are offered with concentration stated, for compounds that do not survive as solids.
Frequently asked
Does the internal standard need a certificate?
How long should the relaxation delay be?
Why does spinning matter?
Sources and further reading
Every document below is published by the body that owns it; none of them is a summary written by us. Accessed 2026-11-24.
- NIST — Standard Reference Materials — certified reference materials and the traceability chain they carry
- ICH — Q2(R2) analytical validation — validation expectations that apply to quantitative procedures
- USP — Reference Standards — how official reference standards are established and documented
Send us a CAS number
Availability, documentation and lead time — answered by the people who handle the batch, not a ticket queue.