A residual solvent limit is a dose calculation, and the table is being revised

Impurities & regulation Published By RSSE Technical Team · Reviewed by RSSE Quality Unit

Table 2 of ICH Q3C gives every Class 2 solvent two numbers, and the ppm column is the PDE divided by a 10 g dose. That division decides what a specification means, why a component can fail while the product passes, and why two entries in the table are under review.

On this page

The ppm column is a division, not a property of the solvent

Table 2 of ICH Q3C(R9) lists 31 Class 2 solvents, and each row carries two numbers: a permitted daily exposure in mg/day and a concentration limit in ppm. The second number is not independent of the first. Section 3.3 of the guideline gives the arithmetic as concentration in ppm equal to 1000 times the PDE divided by the daily dose in grams, and states that the tabulated values were produced by setting the dose to 10 g. Acetonitrile is 4.1 mg/day, and 4.1 divided by 10 and multiplied by 1000 gives 410 ppm. Methanol is 30.0 and 3000. Toluene is 8.9 and 890. Every row obeys the same division, and the same arithmetic for elemental impurities starts from the same kind of table.

So the ppm column answers a question about a 10 g dose. It is not a constant attached to the molecule, and the guideline is explicit that products administered above 10 g per day should be considered under Option 2 instead. A 20 g dose at 410 ppm delivers 8.2 mg of acetonitrile per day, which is twice the PDE.

SolventPDE (mg/day)Table 2 value (ppm, at 10 g/day)At 5 g/dayAt 20 g/day
Acetonitrile4.1410820205
Toluene8.98901780445
Methanol30.0300060001500
Chloroform0.66012030

The rounded figures carry a second warning. The guideline states that PDEs are given to the nearest 0.1 mg/day and concentrations to the nearest 10 ppm, and that the stated values do not reflect the analytical precision of the determination, which is to be established during method validation. A specification written at exactly 410 ppm therefore sits on a rounding boundary and says nothing about whether the method can distinguish 410 from 430.

Option 2 adds exposure across components, so a component can fail while the product passes

Option 1 is a set of default concentrations that may be applied when the daily dose is unknown or fixed, and the guideline notes that components meeting it may be used in any proportion, with no further calculation, provided the dose does not exceed 10 g. Option 2 replaces the default with the actual dose and adds one obligation that is easy to skip: the residual solvent present in every component is summed, and the sum per day, not each component separately, must stay below the PDE. The guideline also requires that the solvent has been reduced to the practical minimum and that the limit is realistic against manufacturing capability.

Its own worked example uses acetonitrile in a 5.0 g product. The Option 1 limit is 410 ppm, but at 5.0 g/day the Option 2 concentration works out at 820 ppm, and only one of the three components complies with 410 ppm.

ComponentMass in the formulationAcetonitrileDaily exposure
Drug substance0.3 g800 ppm0.24 mg
Excipient 10.9 g400 ppm0.36 mg
Excipient 23.8 g800 ppm3.04 mg
Drug product5.0 g728 ppm3.64 mg

The product is at 728 ppm, well above the 410 ppm that a buyer might write into a purchase specification, and it conforms, because 3.64 mg is below 4.1 mg. This is where supply-chain arguments start. An excipient lot at 500 ppm fails a 410 ppm incoming specification and is still perfectly usable in a product dosed at 5 g. The reverse error is more dangerous: a file of conforming certificates of analysis, each component below 410 ppm, does not prove the product conforms, because two components at 400 ppm in 2.5 g each already deliver 2.0 mg, and two at 800 ppm deliver 4.0 mg, which passes by 0.1 mg. The addition has to be done at the product level with the real masses.

Class 3 is a GMP boundary, not a toxicity threshold

Class 3 solvents are defined as those with low toxic potential, for which no health-based exposure limit is needed, and the guideline sets their PDE at 50 mg or more per day. The familiar 5000 ppm, or 0.5 per cent, is that 50 mg divided by 10 g. Amounts above it may still be acceptable where they are realistic for the manufacturing process, so a Class 3 result above 0.5 per cent is a question about the process rather than an automatic failure.

Class 1 sits at the other end of the same table. Benzene is limited to 2 ppm. 1,1,1-Trichloroethane is limited to 1500 ppm in the same table, and the guideline explains why: it is listed because of its environmental effect, and the 1500 ppm figure comes from a review of safety data rather than from the carcinogenicity route that produces the 2 ppm for benzene. A 750-fold spread inside one table means the class number does not predict the order of magnitude.

The table is a moving target, and two entries are moving now

Four of the 31 Class 2 entries have already been revised through the maintenance procedure: N-methylpyrrolidone in September 2002, cumene in February 2011, methylisobutyl ketone in November 2016, and cyclopentyl methyl ether together with tert-butyl alcohol in April 2021. The current revision, Q3C(R9), reached Step 4 on 24 January 2024. The concept paper for Q3C(R10), dated 26 November 2024 and endorsed in January 2025, names two solvents whose values are under review for reasons that have nothing to do with new manufacturing data.

Dichloromethane is Class 2 today at a PDE of 6.0 mg/day, which is 600 ppm at 10 g. The concept paper records that ECHA has proposed reclassifying it from Carc 2 to Carc 1B and adding a Muta 2 classification, which would make it a mutagenic carcinogen, and that the working group will evaluate whether the monograph and the classification in Q3C need revision. If dichloromethane moves to Class 1, it enters a table whose limits are 2, 4, 5 and 8 ppm, and the fall from 600 ppm is between 75-fold and 300-fold. N,N-Dimethylformamide, at 8.8 mg/day and 880 ppm, is under review because IARC moved it from group 2B to 2A.

A limit copied from the table is a snapshot. A purchase agreement naming 600 ppm for dichloromethane, with no clause tying the figure to the current revision, will be wrong in one direction or the other; the direction that matters is the one where the supplier still reports 590 ppm and the method has to reach 8.

What this changes about the standard you buy

A residual solvent standard cannot be prepared in the solvent it measures. A toluene standard is not made in toluene, and the choice of diluent is a technical decision with three checks behind it.

The first is that the diluent must be free of every analyte in scope at the concentration you need to report. Methanol is a Class 2 solvent with a 3000 ppm limit, and a toluene standard in methanol is unusable in a method that also reports methanol, because the diluent peak arrives at the same retention time as the analyte and at a concentration thousands of times higher. Water avoids that for many volatiles but brings a broad solvent peak that tails over the early eluters. The certificate should name the diluent, and the laboratory should confirm the diluent's own analyte content rather than assume that a chromatographic grade solvent is clean at the part-per-billion level.

The second check is matrix. In headspace gas chromatography the peak area depends on the partition coefficient between the sample phase and the gas phase, and that coefficient depends on the matrix, the phase ratio and the thermostat temperature. A standard made up in dimethyl sulfoxide and a sample dissolved in water, at the same concentration, do not give the same area, and the difference is large enough to change a pass into a fail. Either the calibration standards are matrix-matched to the sample, or the result is produced by standard addition, or the method injects liquid and accepts the loss of sensitivity.

The third is quantitation. Benzene is limited to 2 ppm, so a procedure that reports a number against that limit needs a quantitation limit at or below 2 ppm, which in practice means several times lower than the figure being judged. ICH Q2(R2) separates the two cases: for a quantitative impurity test the lower limit of the range is the quantitation limit, while a limit test is supported by a detection limit. A limit test can therefore prove that benzene is absent at 2 ppm without ever producing a number, and a certificate that reports a numerical result has to come from the first kind of procedure.

Storage closes the loop. A volatile standard in a septum-sealed vial loses concentration with every puncture, and a solvent with a boiling point near 40 degrees Celsius is the worst case. Single-use aliquots, or an ampoule opened once, remove the drift; a 10 mL vial used twenty times does not. The hold time a laboratory assigns to its own working solution has the same shape of problem.

A buyer's checklist

  • Ask which solvent, at what concentration, and on which basis: mg/mL for a solution, mg/g for a gravimetric preparation. The two differ by the density.
  • Ask for the diluent by name, and for its analyte content.
  • Ask how the value was assigned, gravimetrically or volumetrically. Dispensing a few microlitres of a solvent with a boiling point of 40 degrees Celsius loses mass between the pipette tip and the vial, and a gravimetric preparation with the mass recorded after sealing does not.
  • Ask for the uncertainty on the assigned value and the coverage factor. Without it, a difference of two per cent between two suppliers cannot be interpreted.
  • Ask whether the standard is a single analyte or a mixture, and if a mixture, which pairs are known to co-elute on the column you intend to use.
  • Ask for the storage condition and the shelf life after the first opening, separately from the shelf life of the unopened container.

A residual solvent limit is an exposure calculation wearing the clothes of a concentration, and the standard is only the ruler used to check it. Getting the arithmetic right at the product level, and the matrix right at the bench, decides whether a compliant batch is rejected or a non-compliant one released.

Frequently asked

Is 410 ppm for acetonitrile a fixed legal limit?
No. It is the PDE of 4.1 mg/day divided by a 10 g daily dose. Option 2 recalculates it from the actual dose: 820 ppm at 5 g per day, 205 ppm at 20 g.
Can a component exceed the Option 1 limit and still be used?
Yes, if the summed daily exposure from all components stays below the PDE and the solvent has been reduced to the practical minimum. The guideline example has components at 800 ppm in a product that conforms.
Does a Class 3 result above 0.5 per cent fail the specification?
Not automatically. The 5000 ppm figure is 50 mg per day at a 10 g dose, and higher amounts may be acceptable where they are realistic for the manufacturing process.
Why is 1,1,1-trichloroethane allowed at 1500 ppm when benzene is limited to 2 ppm?
Because it is listed in Class 1 for its environmental effect, and the 1500 ppm figure comes from a review of safety data. Benzene is limited through the carcinogenicity route.
Is the 600 ppm limit for dichloromethane still current?
Under Q3C(R9) it is. The Q3C(R10) concept paper records an ECHA proposal to reclassify dichloromethane as Carc 1B with Muta 2, and a review of the monograph and of the classification. If it becomes Class 1, the comparable entries in that table are 2 to 8 ppm.
Which diluent should a residual solvent standard use?
One that contains none of the analytes in scope at the reporting level and does not co-elute with them. The certificate should name it, because a diluent that is itself an analyte, such as methanol in a methanol method, makes the standard unusable.

Sources and further reading

Every document below is published by the body that owns it; none of them is a summary written by us. Accessed 2026-11-27.

Send us a CAS number

Availability, documentation and lead time — answered by the people who handle the batch, not a ticket queue.

Request a quote sales@rssechem.com

How to read a certificate of analysis Field by field: identity, purity, content basis, the retest date, and the questions worth asking a supplier before you file the document.How to choose a reference standard supplier A practical checklist for evaluating reference standard suppliers on documentation, traceability, lot consistency, logistics and technical support.Stable isotope-labelled standards and when to use them When labelled internal standards earn their place in an LC-MS or GC-MS method, what to specify when ordering, and the traps in quantitative work.Nitrosamine and NDSRI standards: what to test for and what to buy The difference between small nitrosamines and NDSRIs, what a laboratory needs as a standard for each, and how to specify one so the result stands up to review.What is an impurity reference standard? What the term means, how it differs from a pharmacopoeial standard and a working standard, what the documentation has to state, and how to pick the right form and purity basis.ICH M7 in practice: classifying mutagenic impurities How the five-class hazard assessment works, what the 1.5 microgram default really covers, and how the four control options decide which standards you buy.qNMR vs mass balance: how purity is really assigned Two certificates, one compound, two purity values. What mass balance and qNMR each measure, why the figures differ, and how the basis changes the number.Elemental impurities: from a PDE to a limit you can measure How a permitted daily exposure becomes a concentration limit, why the dose moves the number, and what an elemental impurity specification must state.Salt, hydrate and free base: converting content so the numbers reconcile How to move content and impurity values between as-is, anhydrous and free-base bases, in what order, and why the basis can turn a pass into a fail.Forced degradation study design: how far to push each stressor How to set stressor strength, read a kinetics screen, judge mass balance and catch over-degradation before it wastes a method development cycle.Analytical method transfer: acceptance criteria you can defend Four transfer routes, criteria built from the method's own precision, and the standard's assigned content that shifts every result by a fixed amount.Chiral and stereoisomeric impurity standards Enantiomer, diastereomer, epimer, E/Z isomer: four different standards and four different questions. What each has to state before it can control a method.Reference standard, analytical standard, CRM: what the words mean Four labels appear on the same vial in different catalogues. They are not synonyms, and the difference decides what you may claim in a report.Peptide reference standards: what changes when the analyte is a chain Deletion sequences, epimers, pyroglutamate, deamidation, counter-ion and water: why a peptide standard needs a different certificate from a small molecule.Cleaning validation residue limits: where the number comes from How a residue limit is derived from the dose, from 10 ppm and from a health-based PDE, and what the swab method and its standard must be able to reach.Cold chain and stability of reference standards: what transport and receipt checks protect Storage regimes, stability budgets and mean kinetic temperature, plus the receipt checks that decide whether a standard still supports its certificate.Impurity profile comparison: what a reference standard proves, and what it does not How reporting, identification and qualification thresholds set the scope of an impurity profile comparison, and where a standard stops proving anything.Weighing a reference standard: minimum weight and moisture How the 0.10% repeatability rule sets a minimum weight, why a hygroscopic standard is used as is, and which figures belong on the worksheet.Standard solution stability: how long a prepared standard can be used How to set a hold time for a standard solution: two baselines, six failure modes, a defensible study design and criteria fixed before the data arrive.Reporting, identification and qualification thresholds by dose How the 0.05, 0.10 and 0.15 per cent thresholds are banded by dose, where the milligram cap takes over, and what each limit forces you to test.USP 233 in practice: the J value and the four preparation routes How chapter 233 turns a limit into a J value, which of four preparation routes to use, and the acceptance criteria a method must meet before routine use.Building an uncertainty budget for a standard solution A worked budget for a 1 mg/mL standard solution: weighing, assigned value and volume, the correlation trap, and when k = 2 understates coverage.When the reference standard is the suspect in an OOS result How to test whether a reference standard caused an out-of-specification result, and how to separate a changed material from a misapplied assigned value.Two certificates for one compound: reconciling them field by field How to compare two certificates for the same compound: identity, measurand, basis, assignment, uncertainty, and the order that makes the numbers comparable.How much evidence identifies an unknown impurity How much evidence actually identifies an unknown impurity: formula searches, isotope arithmetic, the limits of fragments, and what a reference standard adds.Relative response factors: what the ratio does to a reported impurity What a relative response factor does to a reported impurity result, why the direction of the bias decides whether it can be waived, and how to measure it.Nitroso impurity numbering: why 3 is not the same compound twice A catalogue number is not a chemical name. How to specify a nitroso impurity by position and structure, before a purchase turns into the wrong isomer.How nitroso impurities form: the two partners and where they meet Nitrosation needs an amine and a nitrosating species in the same place at once. The practical question is where that happens in a process or a formulation.Testing for nitroso impurities: detectors, artefacts and proof Why a standard HPLC-UV method cannot carry a nitrosamine limit, which detectors fit, and what a spiked standard adds that a retention time does not.What a nitroso impurity certificate has to state Identity, assigned content and stability are three separate claims. On a reactive impurity they cannot be copied from the template used for a stable solid.qNMR sample preparation: the steps that decide the number Weighing, solvent, internal standard and acquisition parameters. Where accuracy is won or lost before the first scan is acquired.qNMR calculation: one worked example, every term explained Integral ratio, proton counts, molar masses and the purity of the standard. Where each term comes from, and how a small weighing error moves the answer.The qNMR balance check: reconciling three purity numbers Three numbers describe one powder. When they disagree, the gap is the information: water, residual solvent, inorganic residue or a detector that cannot see.Choosing a qNMR internal standard: spectrum, solubility, stability A single sharp signal is the entry requirement. Solubility, volatility and hygroscopicity decide whether the number survives the weighing.SDS and certificate of analysis: which document answers what The safety data sheet describes hazards and handling. The certificate describes one lot. Audits, shipping papers and storage decisions need the right one.Safety data sheets for stable isotope compounds Stable isotopes are not radioactive, and their SDS should not read as if they were. What changes on the document, and what to check before filing it.What to put in an impurity standard enquiry Most first enquiries cannot be quoted because one field is missing. A short list of what a supplier needs, and why each item changes the answer.Which impurity standards a new API needs first Route-based suspects, degradation products and the ones a method cannot be validated without. A sequence that avoids buying standards twice.System suitability, calibration and reference standards in HPLC Suitability checks the system, calibration checks the response, a standard anchors identity. Mixing them is how a method passes and a result is wrong.GC-MS standards for volatile and semi-volatile impurities Volatility, thermal stability and derivatisation decide the route. What changes in the standard when the detector is a mass spectrometer.Deuterium or carbon-13: choosing the label for the job Cost, exchange, isotope effects and mass resolution. The label choice follows from whether the material is an MS internal standard or an NMR reference.Internal standard correction in ICP-OES and ICP-MS Drift, matrix effects and transport efficiency are different problems. One added element corrects some of them, and the wrong choice hides the rest.A checklist for a custom impurity synthesis enquiry Structure, stereochemistry, quantity, purity target and documentation. Missing any one of them turns a quotation into a correspondence.Pharmacopoeial impurity letters: what A, B and C actually mean Letters are assigned per monograph and can be reassigned when the text is revised. How to map a letter to a structure before you buy it.In-house impurity codes and how to make them portable Company numbering schemes are not transferable. How to convert an internal designation into a specification a supplier can quote against.Metabolite standards: what to specify beyond the structure Hydroxylation position, conjugation, stability in solution and the difference between a bioanalytical standard and a drug substance impurity.